BASELINE
Baseline
position is the position of the wave at the x-axis and y-axis ( 0,0 ). The
graph can only have ±5 mm error from the baseline and this baseline based on
the first graph (at the left side). When measured the baseline position at the
graph paper it should consist only ±5 mm.
Result from conventional ECG machine |
Both of graph
show an acceptable value of error which is below ±5 mm. However this project
contain much higher baseline error value of ±2.45 mm to be compared with
baseline error value of ±0.2 mm for the
conventional ECG machine. Hence,
the baseline condition for the conventional ECG is much more accurate, stable
and acceptable. This problem should be utilise to increase to have a precise
baseline by setting an accurate and correct range of amplitude of Y axis and
time for X-axis. It can be done with the use of properties button in the
LabVIEW software.
For linearity based on the print out graph either it in linear or not and the result is still acceptable if the is only have ±5 % error. Picture below show the linearity of the ECG machine which measured by a straight blue line as an indicator.
Result obtain of this project |
Result from conventional ECG machine |
Both of the
waveform shown a different percentage of linearity which is the conventional
ECG linearity is much more stable and accurate to be compared with the project
linearity. As a solution to this problem, a lower gain should be applied to
this project to ensure its stability and accuracy.